The AP031 is a low cost, battery operated active differential probe intended for measuring higher voltages. The differential techniques employed permit measurements to be taken at two points in a circuit without reference to the ground, allowing the oscilloscope to be safely grounded without the use of opto-isolators or isolating transformers.
The AP031 is a low cost, battery operated active differential probe intended for measuring higher voltages. The differential techniques employed permit measurements to be taken at two points in a circuit without reference to the ground, allowing the oscilloscope to be safely grounded without the use of opto-isolators or isolating transformers.
Discontinued!
This product has been discontinued and is no longer available. There is no direct replacement. Don't worry, we have some options to help you find what you need:
Visit our Test Probes category for similar products.
Or let's do a search and help you find what you need:
This probe is specifically designed for situations where:
Bandwidth | 25 MHz |
Rise Time | 14 ns |
Attenuation | 1:10/1:100 |
Attenuation Accuracy | ±2 % |
Input Resistance | 4 Mω |
Input Capacitance | 5.5 pF each side to ground |
Input Configuration | Differential |
Input Voltage | Maximum Differential 1:100 Range: ±700 V (DC + AC pk) and 700 V rms Maximum Differential 1:10 Range: ±70 V (DC + AC pk) and 70 V rms Maximum Common Mode: ±700 V (DC + AC pk) and 700 V rms Maximum Absolute: ±1400 V ( DC + AC pk) and 1000 V rms |
CMRR | 50 Hz to 80 dB; 20 kHz to 60 dB |
Typical Output Offset | <±5 mV |
Typical Output Noise | 0.7 mV rms typical |
Maximum Input Voltage | 1000 V rms (1400 V DC + peak AC) |
Installation (Overvoltage) | Category III |
Pollution Degree | 2 |
Cable Length | 1.3 m |
Weight (probe only) | 42 g |
Bandwidth | 25 MHz |
Rise Time | 14 ns |
Attenuation | 1:10/1:100 |
Attenuation Accuracy | ±2 % |
Input Resistance | 4 Mω |
Input Capacitance | 5.5 pF each side to ground |
Input Configuration | Differential |
Input Voltage | Maximum Differential 1:100 Range: ±700 V (DC + AC pk) and 700 V rms Maximum Differential 1:10 Range: ±70 V (DC + AC pk) and 70 V rms Maximum Common Mode: ±700 V (DC + AC pk) and 700 V rms Maximum Absolute: ±1400 V ( DC + AC pk) and 1000 V rms |
CMRR | 50 Hz to 80 dB; 20 kHz to 60 dB |
Typical Output Offset | <±5 mV |
Typical Output Noise | 0.7 mV rms typical |
Maximum Input Voltage | 1000 V rms (1400 V DC + peak AC) |
Installation (Overvoltage) | Category III |
Pollution Degree | 2 |
Cable Length | 1.3 m |
Weight (probe only) | 42 g |