Rated at ±0.08% BSL accuracy, this all-stainless-steel transducer measures 0 to 750 psi sealed gauge pressure and provides a 10 mV/V ratiometric output through a twist-lock connector. Its silicon sensor includes a 5-point NIST certificate. It is used for automotive and aerospace testing and OEM test-stand integration.
Rated at ±0.08% BSL accuracy, this all-stainless-steel transducer measures 0 to 750 psi sealed gauge pressure and provides a 10 mV/V ratiometric output through a twist-lock connector. Its silicon sensor includes a 5-point NIST certificate. It is used for automotive and aerospace testing and OEM test-stand integration.
OMEGA piezoresistive pressure transducers have a proven record in high performance commercial and aerospace applications for over 25 years. The PX409 Series uses a highly stable silicon wafer which is micro-machined to precision tolerances and then has silicon strain gauges molecularly embedded into it.
At our state-of-the-art facilities, automated test equipment performs pressure and temperature cycling on 100% of the PX409 transducers. The transducers are then calibrated using extremely high accuracy equipment and a 5-point NIST traceable calibration certificate that is included with each transducer.
Features
Applications
| Series | PX409 Series |
| Type | High accuracy pressure transducer |
| Pressure type | Sealed Gauge |
| Pressure range | 0 to 750 psi (0 to 51.7 bar) |
| Accuracy | ±0.08% BSL maximum (combined linearity, hysteresis and repeatability) |
| Zero balance | ±0.5% FS typical, 1% maximum (1% typical, 2% maximum for 2.5 psi and below) |
| Span setting | ±0.5% FS typical, 1% maximum (1% typical, 2% maximum for 2.5 psi and below); calibrated in the vertical direction with fitting down |
| Output | 10 mV/V |
| Supply voltage | 10 V DC (5 mA), ratiometric 5 to 10 V DC |
| Input/output resistance | 5000 Ω ±20% typical |
| Electrical connection | Twist-Lock (MIL-26482-I 10-6P) |
| Compensated temperature | -20 to 185°F (-29 to 85°C) |
| Operating temperature | -49 to 250°F (-45 to 121°C) |
| Thermal effect on zero | ±0.5% span over the compensated range |
| Thermal effect on span | ±0.5% span over the compensated range |
| Proof (overpressure) pressure | 4 times span |
| Burst / secondary containment pressure | to 3000 psi (207 bar) |
| Long-term stability (1 year) | ±0.1% of full scale (typical) |
| Bandwidth | DC to 1 kHz (typical) |
| Response time | <1 millisecond |
| Pressure cycles | 1 million minimum |
| Shock | 50 g, 11 ms half-sine, vertical and horizontal axis |
| Vibration | 5-2000-5 Hz, 30-minute cycle, Curve L, MIL-STD-810 figure 514-2-2, vertical and horizontal axis |
| Minimum isolation (case to output) | 100 MΩ at 50 V DC |
| Wetted parts | 316 stainless steel |
| Sensing element | Micro-machined silicon wafer with molecularly embedded strain gauges |
| Pressure port | 1/4" NPT Male |
| Enclosure rating | IP65 |
| Protection | Reverse polarity, EMC and power-supply fluctuation protection standard |
| Compliance | CE (emissions IEC 55022 Class B; immunity IEC 61000-4-2, -3, -4, -5, -6, -8), RoHS |
| Calibration | 5-point NIST-traceable certificate included |
| Weight | 4 to 7 oz (115 to 200 g) depending on configuration |
| Series | PX409 Series |
| Type | High accuracy pressure transducer |
| Pressure type | Sealed Gauge |
| Pressure range | 0 to 750 psi (0 to 51.7 bar) |
| Accuracy | ±0.08% BSL maximum (combined linearity, hysteresis and repeatability) |
| Zero balance | ±0.5% FS typical, 1% maximum (1% typical, 2% maximum for 2.5 psi and below) |
| Span setting | ±0.5% FS typical, 1% maximum (1% typical, 2% maximum for 2.5 psi and below); calibrated in the vertical direction with fitting down |
| Output | 10 mV/V |
| Supply voltage | 10 V DC (5 mA), ratiometric 5 to 10 V DC |
| Input/output resistance | 5000 Ω ±20% typical |
| Electrical connection | Twist-Lock (MIL-26482-I 10-6P) |
| Compensated temperature | -20 to 185°F (-29 to 85°C) |
| Operating temperature | -49 to 250°F (-45 to 121°C) |
| Thermal effect on zero | ±0.5% span over the compensated range |
| Thermal effect on span | ±0.5% span over the compensated range |
| Proof (overpressure) pressure | 4 times span |
| Burst / secondary containment pressure | to 3000 psi (207 bar) |
| Long-term stability (1 year) | ±0.1% of full scale (typical) |
| Bandwidth | DC to 1 kHz (typical) |
| Response time | <1 millisecond |
| Pressure cycles | 1 million minimum |
| Shock | 50 g, 11 ms half-sine, vertical and horizontal axis |
| Vibration | 5-2000-5 Hz, 30-minute cycle, Curve L, MIL-STD-810 figure 514-2-2, vertical and horizontal axis |
| Minimum isolation (case to output) | 100 MΩ at 50 V DC |
| Wetted parts | 316 stainless steel |
| Sensing element | Micro-machined silicon wafer with molecularly embedded strain gauges |
| Pressure port | 1/4" NPT Male |
| Enclosure rating | IP65 |
| Protection | Reverse polarity, EMC and power-supply fluctuation protection standard |
| Compliance | CE (emissions IEC 55022 Class B; immunity IEC 61000-4-2, -3, -4, -5, -6, -8), RoHS |
| Calibration | 5-point NIST-traceable certificate included |
| Weight | 4 to 7 oz (115 to 200 g) depending on configuration |