This pH electrode is configured for High Temperature, Insertion Type, 0 to 14 pH. It has ph range of 0 to 14 pH. It suits high-temperature environment and harsh conditions.
This pH electrode is configured for High Temperature, Insertion Type, 0 to 14 pH. It has ph range of 0 to 14 pH. It suits high-temperature environment and harsh conditions.
The high temperature insertion type electrode is designed for periodic exposure to steam sterilization or continuous high temperature. This pH electrode is housed in a durable thermoplastic (PAS) body and is rated for temperatures up to 275°F (275°F (135°C)). The use of the double Porous PTFE liquid junctions with matched viscosity electrolytes provides a reference cell which permits extended periods of pH measurements in the presence of sulfides or other silver complexing agents.
Features
Applications
| Series | PHE-5431-10 Series |
| Type | pH electrode |
| Configuration | High Temperature |
| Configuration | Insertion Type |
| Measurement range | 0 pH to 14 pH |
| Configuration | BNC |
| Configuration | PTFE |
| Material | Glass |
| Configuration | Polymer |
| Configuration | EPR |
| Configuration | 500 psi (34.47 bar) |
| Range | 275°F (135°C) |
| pH Range | 0 to 14 pH |
| Temperature Range | 23 to 275°F (-5 to 135°C) @ 25 psig (1.72 bar) |
| Pressure Range | 500 psig (34.47 bar) @ 77°F (25°C) |
| Accuracy | ±0.1% over full range |
| Impedance | 125 Ω (standard version) |
| Reference Cell | Double junction, KCl/AgCl, KNO3 |
| Reference Junction | Porous PTFE |
| Wetted Materials | PTFE, glass membrane, polymer outer body, EPR "O" Rings |
| Response Time | 10 seconds to 95% of reading |
| Drift | Less than 2 mV per week |
| Body | Durable thermoplastic (PAS) body |
| Connector / Cable | Ten foot cable length and BNC connector are standard |
| Optional ATC Sensor | "-PT100" (254 cm) for 100 Ω Pt RTD or "-PT1K" for 1000 Ω Pt RTD |
| Dimensions | 0.47 x 5" (12 x 127 mm) |
| Series | PHE-5431-10 Series |
| Type | pH electrode |
| Configuration | High Temperature |
| Configuration | Insertion Type |
| Measurement range | 0 pH to 14 pH |
| Configuration | BNC |
| Configuration | PTFE |
| Material | Glass |
| Configuration | Polymer |
| Configuration | EPR |
| Configuration | 500 psi (34.47 bar) |
| Range | 275°F (135°C) |
| pH Range | 0 to 14 pH |
| Temperature Range | 23 to 275°F (-5 to 135°C) @ 25 psig (1.72 bar) |
| Pressure Range | 500 psig (34.47 bar) @ 77°F (25°C) |
| Accuracy | ±0.1% over full range |
| Impedance | 125 Ω (standard version) |
| Reference Cell | Double junction, KCl/AgCl, KNO3 |
| Reference Junction | Porous PTFE |
| Wetted Materials | PTFE, glass membrane, polymer outer body, EPR "O" Rings |
| Response Time | 10 seconds to 95% of reading |
| Drift | Less than 2 mV per week |
| Body | Durable thermoplastic (PAS) body |
| Connector / Cable | Ten foot cable length and BNC connector are standard |
| Optional ATC Sensor | "-PT100" (254 cm) for 100 Ω Pt RTD or "-PT1K" for 1000 Ω Pt RTD |
| Dimensions | 0.47 x 5" (12 x 127 mm) |