Image is for reference only, refer to product info for complete details.
Built for portable temperature measurement, this handheld thermocouple thermometer takes two Type K thermocouple inputs and measures -200 to 1372°C (-328 to 2502°F). It is accurate to ±(0.1% reading + 0.5°C). Powered by 9 V, it works great for portable process temperature checks.
Built for portable temperature measurement, this handheld thermocouple thermometer takes two Type K thermocouple inputs and measures -200 to 1372°C (-328 to 2502°F). It is accurate to ±(0.1% reading + 0.5°C). Powered by 9 V, it works great for portable process temperature checks.
The HH11C and HH12C handheld thermometers feature Type K thermocouple inputs and are available in either single- or dual input models. The HH11C is a single-input model that has auto range resolution of 0.1 or 1°. The HH12C is a dual-input meter that allows the user to display T1, T2, or ∆T (T1 - T2) readings with 0.1 or 1° resolution.
Both models have a 41/2 digit display with a maximum reading of 19999, are °C/°F switchable and feature front-panel offset adjustment and maximum reading/display hold functions. Errors contributed by a particular thermocouple can be practically eliminated, over a limited range, by using the recessed offset adjustment on the face of the meter and an accurate temperature standard.
An LED warning light indicates when an open thermocouple input situation exists. Both units also have improved temperature range and humidity specifications.
Features
Applications
| Display | 4 1/2 digit liquid crystal display (LCD) with a maximum reading of 19999 |
| Low battery indication | Displayed when the battery voltage drops below the operating level |
| Measurement rate | 2.5 times/second |
| Operating environment | 32 to 122°F (0 to 50°C) at <80% RH |
| Storage temperature | -4 to 140°F (-20 to 60°C), 0 to 80% RH with battery removed from meter |
| Accuracy condition | Stated accuracy at 73°F (23°C), <80% RH |
| Battery | 9 V (included) |
| Temperature range | -328 to 2502°F (-200 to 1372°C) |
| Autorange | 0.1°C |
| Temperature accuracy | ±(0.1% rdg + 0.5°C) on -58 to 2502°F (-50 to 1372°C); ±(0.1% rdg + 2°C) on -58 to -328°F (-50 to -200°C) Type K (Chromic-Alum) |
| Input | Dual input (T1, T2, T1-T2) |
| Dimensions | 7.3 x 2.6 x 1.4" (185 H x 65 W x 35 mm D) |
| Weight | 8.5 oz (240 g), including batteries |
| Display | 4 1/2 digit liquid crystal display (LCD) with a maximum reading of 19999 |
| Low battery indication | Displayed when the battery voltage drops below the operating level |
| Measurement rate | 2.5 times/second |
| Operating environment | 32 to 122°F (0 to 50°C) at <80% RH |
| Storage temperature | -4 to 140°F (-20 to 60°C), 0 to 80% RH with battery removed from meter |
| Accuracy condition | Stated accuracy at 73°F (23°C), <80% RH |
| Battery | 9 V (included) |
| Temperature range | -328 to 2502°F (-200 to 1372°C) |
| Autorange | 0.1°C |
| Temperature accuracy | ±(0.1% rdg + 0.5°C) on -58 to 2502°F (-50 to 1372°C); ±(0.1% rdg + 2°C) on -58 to -328°F (-50 to -200°C) Type K (Chromic-Alum) |
| Input | Dual input (T1, T2, T1-T2) |
| Dimensions | 7.3 x 2.6 x 1.4" (185 H x 65 W x 35 mm D) |
| Weight | 8.5 oz (240 g), including batteries |