Omega CNIS16D53-EIT PID Temperature and Process Controller, 1/16 DIN, EtherNet/IP, Dual Display

Model: CNIS16D53-EIT

Omega CNIS16D53-EIT PID Temperature and Process Controller, 1/16 DIN, EtherNet/IP, Dual Display-

Image is for reference only, refer to product info for complete details.

Omega CNIS16D53-EIT PID Temperature and Process Controller, 1/16 DIN, EtherNet/IP, Dual Display

Model: CNIS16D53-EIT

Built for closed-loop PID control, this 1/16 DIN controller takes universal thermocouple, RTD, analog, and strain-gauge inputs and drives solid-state relay outputs. It operates on an AC supply, reads out on a dual 4-digit color LED display, and connects via EtherNet/IP. Best for oven, furnace, and extruder temperature regulation.

Built for closed-loop PID control, this 1/16 DIN controller takes universal thermocouple, RTD, analog, and strain-gauge inputs and drives solid-state relay outputs. It operates on an AC supply, reads out on a dual 4-digit color LED display, and connects via EtherNet/IP. Best for oven, furnace, and extruder temperature regulation.

Your Price $781.64 USD
Availability 2 to 3 Weeks
Quantity

OMEGA CNIS16D53-EIT Offers

The OMEGA(R) iSeries is a family of microprocessor-based instruments offered in three true DIN sizes. All of the instruments share the same set-up and configuration menu and method of operation, a tremendous time saver for integration of a large system.

The iSeries family includes extremely accurate digital panel meters "DPi" and single loop PID controllers "CNi" that are simple to configure and use, while providing tremendous versatility and a wealth of powerful features. The CNi Series covers a broad selection of transducer and transmitter inputs with 2 input models.

The strain and process instruments (CNiS models) measure inputs from load cells, pressure transducers, and most any strain gage sensor as well as process voltage and current ranges.

This CNiS model supports 4- and 6-wire bridge communications, ratiometric measurements. The CNiS features fast and easy "in process" calibration/scaling of the signal inputs to any engineering units. This model also features 10-point linearization which allows the user to linearize the signal input from extremely nonlinear transducers of all kinds.

The iSeries can control simple manual operation to ON-OFF and full Autotune PID control. (Selectable preset tune, adaptive tune, PID, PI, PD control modes.) The dual control outputs can be configured for a variety of independent control and alarm applications such as heat/heat, heat/cool, heat/alarm, cool/cool, cool/alarm or alarm/alarm.

The ramp to setpoint feature allows the user to define the rate of rise to setpoint, minimizing thermal shock to the load during start-up.

The iSeries are 1/8, 1/16 and 1/32 DIN controllers featuring the big iSeries color-changing display. The digits are twice the size of typical 1/8 DIN panel meters. The iSeries feature the only LED displays that can be programmed to change color between GREEN, AMBER, and RED.

Featuring optional "embedded Internet" the iSeries connect directly to an Ethernet network and transmit data in standard TCP/IP packets, or serve Web pages over a LAN or the Internet. The iSeries are also available with serial communications. The user can select from the pushbutton menu between RS232, RS422, and RS485, with straightforward ASCII commands or MODBUS.

Features

  • NEMA 4 (IP65) front bezel
  • Temperature stability 50 ppm/°C
  • Front removable and plug connectors
  • AC or DC powered units
  • Ratiometric mode for strain gages
  • Programmable digital filter
  • Universal inputs
  • Accuracy at 25°C: 0.03% rdg ±0.03% range process and strain
  • Totally programmable color displays (Visual Alarms)
  • User-friendly, simple to configure
  • Free software
  • Full Autotune PID control
  • Embedded Internet connectivity optional
  • RS232 and RS485 serial communications optional
  • 2 control/alarm outputs

Applications

  • Scaling and linearizing signals from nonlinear transducers with 10-point linearization
  • Closing a control loop on a 4 to 20 mA transmitter signal
  • Strain and load-cell measurement from pressure transducers and bridge sensors
  • Remote monitoring and data logging over an Ethernet or RS485 network

OMEGA CNIS16D53-EIT Specifications

Accuracy at 25°C ±0.5°C temp nominal; 0.03% rdg ±0.03% range process and strain
Resolution 1°/0.1°; 10 µV process
Temperature stability RTD 0.04°C/°C
Temperature stability TC at 25°C (77°F) 0.05°C/°C
Temperature stability process 50 ppm/°C
NMRR 60 dB
CMRR 120 dB
A/D conversion Dual slope
Reading rate 3 samples/s
Digital filter Programmable
Display 4-digit 9-segment LED 0.40" (10.2 mm); i32, i16, i16D, i8DV 0.83" (21 mm); i8 0.40" (10.2 mm) and 0.83" (21 mm); i8DH RED, GREEN, and AMBER programmable colors for process variable, setpoint and temperature units
Input types Thermocouple, RTD, analog voltage, analog current
Thermocouple lead resistance 100 Ω max
Thermocouple types (ITS 90) J, K, T, E, R, S, B, C, N, L (J DIN)
RTD input (ITS 68) 100/500/1000 Ω Pt sensor, 2-, 3- or 4-wire; 0.00385 or 0.00392 curve
Voltage input 0 to 100 mV, 0 to 1 V, 0 to 10 Vdc
Input impedance 10 MΩ for 100 mV; 1 MΩ for 1 or 10 Vdc
Current input 0 to 20 mA (5 Ω load)
Configuration Single-ended
Polarity Unipolar
Step response 0.7 sec for 99.9%
Decimal selection temperature None, 0.1
Decimal selection process None, 0.1, 0.01 or 0.001
Setpoint adjustment -1999 to 9999 counts
Span adjustment 0.001 to 9999 counts
Offset adjustment -1999 to 9999
Excitation (not included with communication) 24 Vdc @ 25 mA (not available for low-power option)
Strain accuracy 0.03% reading
Strain resolution 10/1 µV
Strain temperature stability 50 ppm/°C
Strain voltage input 0 to 100 mVdc, -100 mVdc to 1 Vdc, 0 to 10 Vdc
Linearization points Up to 10
Strain excitation (optional in place of communication) 5 Vdc @ 40 mA; 10 Vdc @ 60 mA
Control action Reverse (heat) or direct (cool)
Control modes Time and amplitude proportional control; selectable manual or auto PID, proportional, proportional with integral, proportional with derivative and anti-reset windup, and on/off
Rate 0 to 399.9 s
Reset 0 to 3999 s
Cycle time 1 to 199 s; set to 0 for on/off
Gain 0.5 to 100% of span; setpoints 1 or 2
Damping 0000 to 0008
Soak 00.00 to 99.59 (HH:MM), or OFF
Ramp to setpoint 00.00 to 99.59 (HH:MM), or OFF
Auto tune Operator initiated from front panel
Relay output 250 Vac or 30 Vdc @ 3 A (resistive load); configurable for on/off, PID and ramp and soak
Output 1 SPDT, can be configured as alarm 1 output
Output 2 SPDT, can be configured as alarm 2 output
SSR output 20 to 265 Vac @ 0.05 to 0.5 A (resistive load); continuous
DC pulse output Non-isolated; 10 Vdc @ 20 mA
Analog output (output 1 only) Non-isolated, proportional 0 to 10 Vdc or 0 to 20 mA; 500 Ω max
Output 3 retransmission current 10 V max @ 20 mA output
Output 3 retransmission voltage 20 mA max for 0 to 10 V output
Ethernet Standards compliance IEEE 802.3 10 Base-T
Supported protocols TCP/IP, ARP, HTTPGET
RS232/RS422/RS485 Selectable from menu; both ASCII and MODBUS protocol selectable from menu; programmable 300 to 19.2 Kb; complete programmable setup capability; program to transmit current display, alarm status, min/max, actual measured input value and status
RS485 addressing Addressable from 0 to 199
Connection Screw terminals
Alarm type Same as output 1 and 2
Alarm operation High/low, above/below, band, latch/unlatch, normally open/normally closed and process/deviation; front panel configurations
Programmable analog output Non-isolated, retransmission 0 to 10 Vdc or 0 to 20 mA, 500 Ω max (output 1 only); accuracy is ±1% of FS when following conditions are satisfied: input is not scaled below 1% of input FS, analog output is not scaled below 3% of output FS
Power 90 to 240 Vac ±10%, 50 to 400 Hz, 110 to 375 Vdc, equivalent voltage
Low voltage power option 24 Vac, 12 to 36 Vdc for DPi/CNi/DPiS/CNiS; 20 to 36 Vdc for dual display, ethernet and isolated analog output from qualified safety approved source
Isolation power to input/output 2300 Vac per 1 minute test
Isolation for low voltage power option 1500 Vac per 1 minute test
Isolation power to relay/SSR output 2300 Vac per 1 minute test
Isolation relay/SSR to relay/SSR output 2300 Vac per 1 minute test
Isolation RS232/485 to input/output 500 Vac per 1 minute test
Environmental conditions all models 32 to 131°F (0 to 55°C) 90% RH non-condensing
Approvals UL, C-UL, UKCA, CE per EN61010-1:2001, FM (temperature units only)
Protection NEMA 4X/Type 4 (IP65) front bezel
Environmental conditions 32 to 122°F (0 to 50°C), 90% RH non-condensing (for UL only)
Dimensions 1.89 x 1.89 x 5" (48 H x 48 W x 127 mm D)
Panel cutout: 1.772" (45 mm) square, 1/16 DIN
Weight 0.35 lb (159 g)

What's included with the OMEGA CNIS16D53-EIT

  • 1/16 DIN Panel Mount PID Controller
  • Operator's Manual

OMEGA CNIS16D53-EIT Offers

The OMEGA(R) iSeries is a family of microprocessor-based instruments offered in three true DIN sizes. All of the instruments share the same set-up and configuration menu and method of operation, a tremendous time saver for integration of a large system.

The iSeries family includes extremely accurate digital panel meters "DPi" and single loop PID controllers "CNi" that are simple to configure and use, while providing tremendous versatility and a wealth of powerful features. The CNi Series covers a broad selection of transducer and transmitter inputs with 2 input models.

The strain and process instruments (CNiS models) measure inputs from load cells, pressure transducers, and most any strain gage sensor as well as process voltage and current ranges.

This CNiS model supports 4- and 6-wire bridge communications, ratiometric measurements. The CNiS features fast and easy "in process" calibration/scaling of the signal inputs to any engineering units. This model also features 10-point linearization which allows the user to linearize the signal input from extremely nonlinear transducers of all kinds.

The iSeries can control simple manual operation to ON-OFF and full Autotune PID control. (Selectable preset tune, adaptive tune, PID, PI, PD control modes.) The dual control outputs can be configured for a variety of independent control and alarm applications such as heat/heat, heat/cool, heat/alarm, cool/cool, cool/alarm or alarm/alarm.

The ramp to setpoint feature allows the user to define the rate of rise to setpoint, minimizing thermal shock to the load during start-up.

The iSeries are 1/8, 1/16 and 1/32 DIN controllers featuring the big iSeries color-changing display. The digits are twice the size of typical 1/8 DIN panel meters. The iSeries feature the only LED displays that can be programmed to change color between GREEN, AMBER, and RED.

Featuring optional "embedded Internet" the iSeries connect directly to an Ethernet network and transmit data in standard TCP/IP packets, or serve Web pages over a LAN or the Internet. The iSeries are also available with serial communications. The user can select from the pushbutton menu between RS232, RS422, and RS485, with straightforward ASCII commands or MODBUS.

Features

  • NEMA 4 (IP65) front bezel
  • Temperature stability 50 ppm/°C
  • Front removable and plug connectors
  • AC or DC powered units
  • Ratiometric mode for strain gages
  • Programmable digital filter
  • Universal inputs
  • Accuracy at 25°C: 0.03% rdg ±0.03% range process and strain
  • Totally programmable color displays (Visual Alarms)
  • User-friendly, simple to configure
  • Free software
  • Full Autotune PID control
  • Embedded Internet connectivity optional
  • RS232 and RS485 serial communications optional
  • 2 control/alarm outputs

Applications

  • Scaling and linearizing signals from nonlinear transducers with 10-point linearization
  • Closing a control loop on a 4 to 20 mA transmitter signal
  • Strain and load-cell measurement from pressure transducers and bridge sensors
  • Remote monitoring and data logging over an Ethernet or RS485 network

OMEGA CNIS16D53-EIT Specifications

Accuracy at 25°C ±0.5°C temp nominal; 0.03% rdg ±0.03% range process and strain
Resolution 1°/0.1°; 10 µV process
Temperature stability RTD 0.04°C/°C
Temperature stability TC at 25°C (77°F) 0.05°C/°C
Temperature stability process 50 ppm/°C
NMRR 60 dB
CMRR 120 dB
A/D conversion Dual slope
Reading rate 3 samples/s
Digital filter Programmable
Display 4-digit 9-segment LED 0.40" (10.2 mm); i32, i16, i16D, i8DV 0.83" (21 mm); i8 0.40" (10.2 mm) and 0.83" (21 mm); i8DH RED, GREEN, and AMBER programmable colors for process variable, setpoint and temperature units
Input types Thermocouple, RTD, analog voltage, analog current
Thermocouple lead resistance 100 Ω max
Thermocouple types (ITS 90) J, K, T, E, R, S, B, C, N, L (J DIN)
RTD input (ITS 68) 100/500/1000 Ω Pt sensor, 2-, 3- or 4-wire; 0.00385 or 0.00392 curve
Voltage input 0 to 100 mV, 0 to 1 V, 0 to 10 Vdc
Input impedance 10 MΩ for 100 mV; 1 MΩ for 1 or 10 Vdc
Current input 0 to 20 mA (5 Ω load)
Configuration Single-ended
Polarity Unipolar
Step response 0.7 sec for 99.9%
Decimal selection temperature None, 0.1
Decimal selection process None, 0.1, 0.01 or 0.001
Setpoint adjustment -1999 to 9999 counts
Span adjustment 0.001 to 9999 counts
Offset adjustment -1999 to 9999
Excitation (not included with communication) 24 Vdc @ 25 mA (not available for low-power option)
Strain accuracy 0.03% reading
Strain resolution 10/1 µV
Strain temperature stability 50 ppm/°C
Strain voltage input 0 to 100 mVdc, -100 mVdc to 1 Vdc, 0 to 10 Vdc
Linearization points Up to 10
Strain excitation (optional in place of communication) 5 Vdc @ 40 mA; 10 Vdc @ 60 mA
Control action Reverse (heat) or direct (cool)
Control modes Time and amplitude proportional control; selectable manual or auto PID, proportional, proportional with integral, proportional with derivative and anti-reset windup, and on/off
Rate 0 to 399.9 s
Reset 0 to 3999 s
Cycle time 1 to 199 s; set to 0 for on/off
Gain 0.5 to 100% of span; setpoints 1 or 2
Damping 0000 to 0008
Soak 00.00 to 99.59 (HH:MM), or OFF
Ramp to setpoint 00.00 to 99.59 (HH:MM), or OFF
Auto tune Operator initiated from front panel
Relay output 250 Vac or 30 Vdc @ 3 A (resistive load); configurable for on/off, PID and ramp and soak
Output 1 SPDT, can be configured as alarm 1 output
Output 2 SPDT, can be configured as alarm 2 output
SSR output 20 to 265 Vac @ 0.05 to 0.5 A (resistive load); continuous
DC pulse output Non-isolated; 10 Vdc @ 20 mA
Analog output (output 1 only) Non-isolated, proportional 0 to 10 Vdc or 0 to 20 mA; 500 Ω max
Output 3 retransmission current 10 V max @ 20 mA output
Output 3 retransmission voltage 20 mA max for 0 to 10 V output
Ethernet Standards compliance IEEE 802.3 10 Base-T
Supported protocols TCP/IP, ARP, HTTPGET
RS232/RS422/RS485 Selectable from menu; both ASCII and MODBUS protocol selectable from menu; programmable 300 to 19.2 Kb; complete programmable setup capability; program to transmit current display, alarm status, min/max, actual measured input value and status
RS485 addressing Addressable from 0 to 199
Connection Screw terminals
Alarm type Same as output 1 and 2
Alarm operation High/low, above/below, band, latch/unlatch, normally open/normally closed and process/deviation; front panel configurations
Programmable analog output Non-isolated, retransmission 0 to 10 Vdc or 0 to 20 mA, 500 Ω max (output 1 only); accuracy is ±1% of FS when following conditions are satisfied: input is not scaled below 1% of input FS, analog output is not scaled below 3% of output FS
Power 90 to 240 Vac ±10%, 50 to 400 Hz, 110 to 375 Vdc, equivalent voltage
Low voltage power option 24 Vac, 12 to 36 Vdc for DPi/CNi/DPiS/CNiS; 20 to 36 Vdc for dual display, ethernet and isolated analog output from qualified safety approved source
Isolation power to input/output 2300 Vac per 1 minute test
Isolation for low voltage power option 1500 Vac per 1 minute test
Isolation power to relay/SSR output 2300 Vac per 1 minute test
Isolation relay/SSR to relay/SSR output 2300 Vac per 1 minute test
Isolation RS232/485 to input/output 500 Vac per 1 minute test
Environmental conditions all models 32 to 131°F (0 to 55°C) 90% RH non-condensing
Approvals UL, C-UL, UKCA, CE per EN61010-1:2001, FM (temperature units only)
Protection NEMA 4X/Type 4 (IP65) front bezel
Environmental conditions 32 to 122°F (0 to 50°C), 90% RH non-condensing (for UL only)
Dimensions 1.89 x 1.89 x 5" (48 H x 48 W x 127 mm D)
Panel cutout: 1.772" (45 mm) square, 1/16 DIN
Weight 0.35 lb (159 g)

What's included with the OMEGA CNIS16D53-EIT

  • 1/16 DIN Panel Mount PID Controller
  • Operator's Manual

More OMEGA Products