Connect this signal source to your device to perform loop testing, maintenance or calibrate thermometers. The compact unit can also be used to measure DC voltage and current.
Connect this signal source to your device to perform loop testing, maintenance or calibrate thermometers. The compact unit can also be used to measure DC voltage and current.
Capable of generating 0 to 25 V and up to 25 mA, this Hioki SS7012 DC Signal Source is an ideal tool for your maintenance or loop testing needs. The unit also boasts 8 selectable thermocouples, so you can calibrate thermometers to check temperature control equipment.
Features
Circuit Method | Bipolar sink and source |
Constant Voltage | 2.5 V: 0 to ±2.5 V (±0.03% of setting ±300 μV, 100 μV resolution), 25 V: 0 to ±25 V (±0.03% of setting ±3 mV, 1 mV resolution) |
Constant Current | 0 to ±25 mA (±0.03% of setting ±3μA, 1μA resolution) |
Thermoelectric Power Generation | K (at TC 32°F (0°C)): -281.2 to 2,501.6°F (-174 to 1,372°C) (±0.05% of setting ±0.5°C, 0.1°C resolution), K (at TC RJ): -281.2 to 2,501.6°F (-174 to 1,372°C) (±0.05% of setting ±1°C, 0.1°C resolution), Other Types: E, J, T, R, S, B, N selectable |
Standard Resistance (Rs) | 100 Ω (±0.2 Ω) |
Automatic Generation | Number of memory steps: 20, Interval time: 1 to 99 seconds (at CV, CC, TC mode) |
Measurement Voltage | 2.5 V: 0 to ±2.8 V (±0.03% of reading ±300 μV, 100 μV, resolution, 1M Ω input resistance), 25 V: 0 to ±28 V (±0.03% of reading ±3μA, 1μA resolution, smaller than 30 Ω input resistance) |
Measurement Current | 0 to ±28 mA (±0.03% of reading ±3μA, 1μA resolution, smaller than 30 Ω input resistance) |
Measurement Temperature | -13 to 176°F (-25 to 80°C) (±0.5°C at 23 ±5°C, 0.1°C resolution, use with the RJ sensor 9184) |
Sampling Rate | Approximately 1.67 times/second |
Other Functions | Zero adjustment, overflow display, USB communication, monitor |
Power Supply | 4 LR6 AA batteries, 9445-02/-03 AC adapter, 4 HR6 NiMH batteries |
Product Length | 2.28" (58 mm) |
Product Width | 4.09" (104 mm) |
Product Height | 7.09" (180 mm) |
Product Weight | 23.3 oz (660 g) |
Circuit Method | Bipolar sink and source |
Constant Voltage | 2.5 V: 0 to ±2.5 V (±0.03% of setting ±300 μV, 100 μV resolution), 25 V: 0 to ±25 V (±0.03% of setting ±3 mV, 1 mV resolution) |
Constant Current | 0 to ±25 mA (±0.03% of setting ±3μA, 1μA resolution) |
Thermoelectric Power Generation | K (at TC 32°F (0°C)): -281.2 to 2,501.6°F (-174 to 1,372°C) (±0.05% of setting ±0.5°C, 0.1°C resolution), K (at TC RJ): -281.2 to 2,501.6°F (-174 to 1,372°C) (±0.05% of setting ±1°C, 0.1°C resolution), Other Types: E, J, T, R, S, B, N selectable |
Standard Resistance (Rs) | 100 Ω (±0.2 Ω) |
Automatic Generation | Number of memory steps: 20, Interval time: 1 to 99 seconds (at CV, CC, TC mode) |
Measurement Voltage | 2.5 V: 0 to ±2.8 V (±0.03% of reading ±300 μV, 100 μV, resolution, 1M Ω input resistance), 25 V: 0 to ±28 V (±0.03% of reading ±3μA, 1μA resolution, smaller than 30 Ω input resistance) |
Measurement Current | 0 to ±28 mA (±0.03% of reading ±3μA, 1μA resolution, smaller than 30 Ω input resistance) |
Measurement Temperature | -13 to 176°F (-25 to 80°C) (±0.5°C at 23 ±5°C, 0.1°C resolution, use with the RJ sensor 9184) |
Sampling Rate | Approximately 1.67 times/second |
Other Functions | Zero adjustment, overflow display, USB communication, monitor |
Power Supply | 4 LR6 AA batteries, 9445-02/-03 AC adapter, 4 HR6 NiMH batteries |
Product Length | 2.28" (58 mm) |
Product Width | 4.09" (104 mm) |
Product Height | 7.09" (180 mm) |
Product Weight | 23.3 oz (660 g) |